Extech Instruments NON-DISPLAY PORTABLE Especificações Página 6

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 13
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 5
CG204-EU-EN v2.2 8/13
6
c. ProbeUsed
SelectAUTO,Fe,orNoFeunderPROBEUSEDintheOPTIONSMenuusingthearrowbuttonsandthe
SELECTsoftkey.
IntheAUTOmode,themeterautomaticallyactivatestheprobemeasurementmethod(ferrousornon
ferrous)basedonthemetalsubstratethatisbeingmeasured.Whentheprobeis
placedonamagnetic
substrateitwillworkinthemagneticinductionmode.Iftheprobeisplacedonanonferrousmetalit
willworkintheeddycurrentmode.
IntheFerrous(Fe)ModetheMagneticinductionmeasurementmodeisactivated.
IntheNonFerrous(NoFe)Mode
theeddycurrentmeasurementmodeisactivated.
d. Unitsofmeasureselections
Selectmm,µm,ormilsunderUNITSETTINGintheOPTIONSMenuusingthearrowbuttonsandthe
SELECTsoftkey(mm=millimeters;µm=micrometers;mils=mm*2.54/100)
e. Backlight
SelectONorOFFunderBACKLIGHTintheOPTIONSMenuusingthearrowbuttonsandtheSELECTsoft
key.IfOFFisselectedtheLCDbacklightingwillbecompletelydisabled.IfONisselected,theusercan
turnthelightonoroffusingthebacklightbutton(CANCELbutton).
f. LCDStatistics
SelectAVERAGE,MINIMUM,MAXIMUM,orSDEV(StandardDeviation)underLCDSTATISTICSinthe
OPTIONSMenuusingthearrowbuttonsandtheSELECTsoftkey.Thisselectiondetermineswhich
statisticisshownasdefaultontheLCDdisplay.
g. AutoPowerOFF
SelectENABLEorDISABLEunderAUTOPOWEROFFintheOPTIONSMenuusingthearrowbuttonsand
theSELECTsoftkey.Whenenabled,themeterautomaticallyswitchesOFFafter3minutesofinactivity.
Whendisabled,themeterwillonlyswitchOFFwithabuttonpressorwhenthebatterypoweris
weak.
LIMITMenu
HighandLowAlarmLimitscanbesetwhenanAlarmLimitisreachedthemeterdisplaysanAlarmAlerticon
(HforHighAlarmandLforLowAlarm).
1. PresstheMENUbuttontoaccesstheprogrammingmenu
2. UsertheDOWNARROWbuttontoscrolldownto
LIMITandpressSELECT
3. PressSELECTagaintochooseLIMITSETTING
4. PressSELECTagaintochooseHIGHLIMIT
5. UsertheARROWBUTTONStosetaHighAlarmvalue
6. PressOKtostorethelimitandpressBACKtoreturntothemenu
7. Performthesamesteps
fortheLOWLIMIT
8. UsetheDELETELIMITparametertoclearAlarmLimitvalues
Vista de página 5
1 2 3 4 5 6 7 8 9 10 11 12 13

Comentários a estes Manuais

Sem comentários