User's Guide
Noise Dosimeter with PC Interface
Model 407355
Introduction
Congratulations on your purchase of the Extech 407355 Personal Noise Dosimeter. The
407355 is designed to test noise exposure in accordance with OSHA, MSHA, DOD,
ACGIH, and ISO standards. Fast and easy on-site surveys help determine noise
reduction requirements. The meter can also be used in SLM (sound level meter) mode
where sound pressure levels are monitored from 70 to 140dB. The SLM (sound level
meter) mode has a datalogging feature that can record up to 8800 readings which can
be downloaded to a PC for analysis. Note: In DOSIMETER mode, the individual
measurements are NOT datalogged and cannot be recalled.
The built-in RS-232 PC interface offers the following capabilities:
In SOUND LEVEL METER (SLM) MODE: The meter can be connected to PC while
taking measurements so that readings can be stored in a file and readings can be
viewed in a variety of display configurations (graph, list, etc.).
In DOSIMETER mode: The meter can perform a noise survey after which the user
can transfer the survey data to a PC.
Careful use of this device will provide years of reliable service.
Vista de página 0
1 2 3 4 5 6 ... 16 17

Resumo do Conteúdo

Página 1 - Model 407355

User's Guide Noise Dosimeter with PC Interface Model 407355 Introduction Congratulations on your purchase of the Extec

Página 2 - Warranty

407355 Version 2.4 9/07 10 Software Reference Information The REMOTE CONTROL and MAIN SOFTWARE Windows When the DOSE program is opened, two windows

Página 3 - Descriptions

407355 Version 2.4 9/07 11 MAIN Software Window Menu Headings FILE OPEN: Use an existing data file or create a new data file to store readings in S

Página 4

407355 Version 2.4 9/07 12 EVENT WINDOW - Download data from meter bank (E1 through E5) to PC In this window the user can download, view, and save t

Página 5 - Operation

407355 Version 2.4 9/07 13 AVERAGE When AVERAGE is selected the window shown in the accompanying diagram opens. In the AVERAGE mode, the user select

Página 6

407355 Version 2.4 9/07 14 OCCUPATIONAL NOISE EXPOSURE STANDARD Window The Occupational Noise Exposure Standard window (see diagram) is a convenient

Página 7 - Manual Programming

407355 Version 2.4 9/07 15 DIN to DB-9 CablePC Interface Cable and Adaptor Wiring Reference The wiring diagrams and pin-outs for the supplied PC I

Página 8 - Calibration Procedure

407355 Version 2.4 9/07 16 Conversion Charts OSHA Permissible Noise Exposures Duration per day, hours Sound level dBA, slow response 8 90 6 92 4 95

Página 9

407355 Version 2.4 9/07 17  Support line (781) 890-7440 Technical support: Extension 200; E-mail: [email protected] Repair & Returns: Extensio

Página 10 - CONTROL PANEL WINDOW

407355 Version 2.4 9/07 2 CCWarranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one

Página 11

407355 Version 2.4 9/07 3 Specifications Display Multi-function LCD Units of measure dBA (‘A’ weighted decibels) for sound level % DOSE for accumul

Página 12

407355 Version 2.4 9/07 4 Meter Description 1. Microphone (lapel clip not shown) 2. Microphone cable 3. Microphone connector 4. LCD display 5. Pus

Página 13

407355 Version 2.4 9/07 5 Operation Getting Started 1. Power the instrument by pressing the green button. 2. When the unit is turned on, all

Página 14

407355 Version 2.4 9/07 6 9. Use the MODE button at any time to view the elapsed time of the test. Press the MODE button until the TIME display ico

Página 15 - DIN to DB-9 Cable

407355 Version 2.4 9/07 7 LCLTER20040_7_0_8_0_3_37SFPEAK FLAG When sound levels above 140dB are present the meter displays the PK symbol HIGH LEVEL

Página 16 - Conversion Charts

407355 Version 2.4 9/07 8 Calibration Procedure Note that an acoustical calibrator (such as the Extech 407744 or 407766) is required to calibrate th

Página 17 - Support line (781) 890-7440

407355 Version 2.4 9/07 9 PC Interface Operational Modes The PC Interface for the Model 407355 is used for both the Sound Level Meter (SLM) and the