9 475040-044-SD-EU-EN V1.5 8/12
Setup Menu
In the Normal mode, press and Hold the SET Button >2 seconds to enter the Setup mode. With
each press of the SET button the programmable functions will appear in this order:
Sd F
SD memory card Format
dAtE
Set clock time ( Year/Month/Date, Hour/Minute/Second )
SP-t
Set sampling time for normal record mode
HSPt
Set sampling time for peak record mode
PoFF
Auto power OFF enable/disable
bEEP
Set beeper sound ON/OFF
unit
Set the measurement unit
dEC
Set SD card Decimal character
ESC
Escape or exit from the setup mode
Sd F - SD Card Formatting
1. Use the arrow buttons to select “yES” to format the card (select “no” to exit).
Note: All data on the card will be lost if formatting is performed.
2. If “yES” is selected, press ENTER to confirm selection and then press ENTER a second time
to confirm the selection.
3. When the formatting is complete, the meter will return to the normal operating mode.
dAtE
- Set Clock Time ( Year/Month/Date, Hour/Minute/ Second )
1. Use the button or button to adjust the flashing digits to the correct setting.
2. When the desired value is set, press the ENTER button and the adjustment will move to the
next value for adjustment.
3. Repeat this for the six values.
4. Press the SET button to save the setting and move to the next function.
Note: The clock data will remain set, even with power off, as long as the batteries are have a
charge.
SP-t -
Set Sampling Time ( normal record mode )
1. Use the button or button to adjust the flashing digits to the desired setting.
2. When the desired value is set, press the ENTER button and the adjustment will move to the
next value for adjustment.
3. Repeat this for the three values.
4. Press the SET button to save the setting and move to the next function.
Note: The default sample rate is 2 seconds.
HSPt
- Set Peak Sampling Time (peak record mode)
1. Use the button or button to adjust the digits to the desired setting. The adjustment range
is 10ms to 500ms in 10ms steps.
2. When the desired value is set, press the ENTER button to store the setting.
3. Press the SET button to move to the next function.
Note: Maximum of 1000 data points can be stored in the Peak mode.