Extech Instruments MG300 Guia do Utilizador Página 18

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 22
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 17
18 MG300-en-US_v2.2 04/15
Meg OHMS
Terminal Voltage Range Resolution Accuracy Test current Short circuit
current
125V (0%~+10%) 0.125~4.000 M 0.001M +(2%+10) 1mA @ load
125k
1mA
4.001~40.00 M 0.01M +(2%+10)
40.01~400.0 M 0.1M +(4%+5)
400.1~4000 M 1M +(5%+5)
250V (0%~+10%) 0.250~4.000 M 0.001M +(2%+10) 1mA @ load
250k
1mA
4.001~40.00 M 0.01M +(2%+10)
40.01~400.0 M 0.1M +(3%+5)
400.1~4000 M 1M +(4%+5)
500V (0%~+10%) 0.500~4.000 M 0.001M +(2%+10) 1mA @ load
500k
1mA
4.001~40.00 M 0.01M +(2%+10)
40.01~400.0 M 0.1M +(2%+5)
400.1~4000 M 1M +(4%+5)
1000V
(0%~+10%)
1.000~4.000 M 0.001M +(3%+10) 1mA @ load
1M
1mA
4.001~40.00 M 0.01M +(2%+10)
40.01~400.0 M 0.1M +(2%+5)
400.1~4000 M 1M +(4%+5)
Note: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.
Vista de página 17
1 2 ... 13 14 15 16 17 18 19 20 21 22

Comentários a estes Manuais

Sem comentários